Guidewires - Endoscopic (Jag Wires) Model #'s: MIC5658 and MIC630-100 Intended to be used for selective...
Sterilmed Inc began a recall of Guidewires - Endoscopic (Jag Wires) Model #'s: MIC5658 and MIC630-100 Intended to be used for selective cannulization of the billary ducts, including but not... on . The reason FDA records is "SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0558-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sterilmed Inc
- Product
- Guidewires - Endoscopic (Jag Wires) Model #'s: MIC5658 and MIC630-100 Intended to be used for selective cannulization of the billary ducts, including but not limited to, the common bile, cystic, right and left hepatic ducts during endoscopic billary procedures for catheter introduction and exchanges.
- Product code
- NML Catheter, Biliary, Reprocessed
- Reason for recall
- SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.
- Quantity in commerce
- 2
- Distribution
- AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA
- Product codes and lots
- n/a
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
