GE Stereotaxy Positioner, model 2405544-2, for use with Senographe DS Full Field Mammography system, models...
GE Healthcare began a recall of GE Stereotaxy Positioner, model 2405544-2, for use with Senographe DS Full Field Mammography system, models 2383168, 2383168-2, 2383168-3, 2383168-3-1... on . The reason FDA records is "GE Healthcare has recently become aware of x-ray emission beyond the edge of the detector primary barrier. This issue occurs when an exam is performed in a specific angulated view associated with use of the Stereotactic...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2146-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Stereotaxy Positioner, model 2405544-2, for use with Senographe DS Full Field Mammography system, models 2383168, 2383168-2, 2383168-3, 2383168-3-1, 2383168-4-1. The expected use of the Senographe DS Stereotaxy is an optional accessory for the Senographe system for mammography examinations.
- Product code
- IZH System, X-Ray, Mammographic
- Reason for recall
- GE Healthcare has recently become aware of x-ray emission beyond the edge of the detector primary barrier. This issue occurs when an exam is performed in a specific angulated view associated with use of the Stereotactic Positioner of your Senographe DS Acquisition system and could impact patient safety. If the Stereotactic Positioner is removed when the examination arm is at 33 degrees and, th
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- A Product Safety Notification letter, dated March 3, 2008, was sent to affected customers. The letter provided information about the safety issue, identified affected product and stated that GE Healthcare will provide a software upgrade. Contact GE Healthcare at 1-262-544-3894 for assistance.
- Quantity in commerce
- 27
- Distribution
- AL, CA, CO, FL, IL, KS, MA, ME, MN,MO, NC, NJ, NM, NY OK, SC, TN, TX, and WI.
- Product codes and lots
- model 2405544-2, serial numbers: 00000434834BU8, 00000434837BU1, 00000437358BU5, 00000437363BU5, 00000437366BU8, 00000437366BU8, 00000444593BU8, 00000444603BU5, 00000450276BU1, 00000450289BU4, 00000454195BU9, 00000454197BU5, 00000454200BU7, 00000454202BU3, 0000046417BU0, 00000464622BU0, 00000464628BU7, 00000484259BU7, 00000484270BU4, 00000484290BU2, 00000484297BU7, 0000048429BU4, 00000492966BU7, 00000492975BU8, 00000494921BU0, 00000500104BU5, and 00000500118BU5.