Z-2147-2008

GE Stereotaxy Positioner, model 2405544-3, for use with Senographe DS Full Field Mammography system, models...

GE Healthcare / initiated 2008-07-01 / Terminated

GE Healthcare began a recall of GE Stereotaxy Positioner, model 2405544-3, for use with Senographe DS Full Field Mammography system, models 2383168, 2383168-2, 2383168-3, 2383168-3-1... on . The reason FDA records is "GE Healthcare has recently become aware of x-ray emission beyond the edge of the detector primary barrier. This issue occurs when an exam is performed in a specific angulated view associated with use of the Stereotactic...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2147-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Stereotaxy Positioner, model 2405544-3, for use with Senographe DS Full Field Mammography system, models 2383168, 2383168-2, 2383168-3, 2383168-3-1, 2383168-4-1. The expected use of the Senographe DS Stereotaxy is an optional accessory for the Senographe system for mammography examinations.
Product code
IZH System, X-Ray, Mammographic
Reason for recall
GE Healthcare has recently become aware of x-ray emission beyond the edge of the detector primary barrier. This issue occurs when an exam is performed in a specific angulated view associated with use of the Stereotactic Positioner of your Senographe DS Acquisition system and could impact patient safety. If the Stereotactic Positioner is removed when the examination arm is at 33 degrees and, th
FDA root cause
Radiation Control for Health and Safety Act
Action
A Product Safety Notification letter, dated March 3, 2008, was sent to affected customers. The letter provided information about the safety issue, identified affected product and stated that GE Healthcare will provide a software upgrade. Contact GE Healthcare at 1-262-544-3894 for assistance.
Quantity in commerce
12
Distribution
AL, CA, CO, FL, IL, KS, MA, ME, MN,MO, NC, NJ, NM, NY OK, SC, TN, TX, and WI.
Product codes and lots
model 2405544-3, Serial numbers: 00000509762BU1, 00000509764BU7, 00000543767BU8, 00000546136BU3, 00000546140BU5, 00000546772BU5, 00000546793BU1, 00000546794BU9, 00000546799BU8, 00000550251BU3, 00000550265BU3, and 00000550266BU1.