Z-0192-2009
Hydrophobic HALO360 Filter that is packaged with HALO360+ Ablation Catheters Model numbers 32041-22...
Medtronic, Inc. / initiated 2008-09-18 / Terminated
Barrx Medical, Inc. began a recall of Hydrophobic HALO360 Filter that is packaged with HALO360+ Ablation Catheters Model numbers 32041-22, 32041-28, 32041-31 and 32041-34, and/or with HALO360... on . The reason FDA records is "Some units may contain the wrong filter, which does not have the proper lock and may result in a leak". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0192-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Hydrophobic HALO360 Filter that is packaged with HALO360+ Ablation Catheters Model numbers 32041-22, 32041-28, 32041-31 and 32041-34, and/or with HALO360 Sizing Balloon Model number 3441B.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Some units may contain the wrong filter, which does not have the proper lock and may result in a leak.
- FDA root cause
- Mixed-up of materials/components
- Action
- The recall was initiated on September 18, 2008. All users and international distributors were informed by letter. The replacement filters were sent on 9/19/2008. Contact Barrx Medical, Inc. at 1-408-745-8000 for assistance.
- Quantity in commerce
- 650 units
- Distribution
- Worldwide Distribution, including USA, Austria, Australia, Belgium, Canada, France, Netherlands, Switzerland, Sweden, and UK .
- Product codes and lots
- Filter Lot number 78156 was packaged with Catheter Lot numbers: F1012714 , F1012715 , F1012716 , F1012718 , F1012721 , F1012723 , and F1012724, and/or with Sizing Balloon Lot number: F1012719.