Z-2118-2013Class II

Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS...

Ventana Medical Systems Inc / initiated 2008-06-12 / Terminated / root cause: design

Ventana Medical Systems Inc began a recall of Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The Benchmark... on . FDA classified it as Class II and gives the reason as "Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2118-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ventana Medical Systems Inc
Product
Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The Benchmark product lines are intended to automatically stain histological or cytological specimens on microscope slides with specific immunohistochemistry or in situ hybridization reagents for in vitro diagnostic use. The Discovery Systems are research use only instruments with similar functionality.
Product code
KPA Slide Stainer, Automated
Reason for recall
Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.
Root cause tag
design
FDA root cause
Device Design
Action
Ventana sent an Urgent Medical Device Correction Notification letter dated June 16, 2008 to all affected customers. The letter identifed the affected product, problem and actions to be taken. The letter instructed customers to carefully read and perform all of the actions outlined in the notification letter concerning potential safety issue. Customers with concerns regarding the staining results on any patient specimen are instructed to follow their internal Quality Procedures regarding the review of patient reports. For question contact your local Customer Service Center.
Quantity in commerce
1217 units total
Distribution
Worldwide Distribution - USA Nationwide and the countries of Canada, Europe, Asia, Japan.
Product codes and lots
BenchMark XT Instruments-S/N: 711869-712465; BenchMark LT instruments-S/N: 610321-610345. Discovery XT Instruments-S/N: 610321-610345.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2118-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2118-2013Class IITerminatedVoluntary: Firm initiated