Z-0113-2009

Zilver Expandable Metal Biliary Stent System, ZILBS-10-6, Stent Diameter: 10mm, Stent Length: 6cm...

Wilson-Cook Medical Inc. / initiated 2008-09-11 / Terminated / root cause: manufacturing process

Cook Endoscopy began a recall of Zilver Expandable Metal Biliary Stent System, ZILBS-10-6, Stent Diameter: 10mm, Stent Length: 6cm, Introduction System: 7 FR., Disposable-Single Use Only, Rx... on . The reason FDA records is "A section of the introduction system may detach after the stent has deployed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0113-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Wilson-Cook Medical Inc.
Product
Zilver Expandable Metal Biliary Stent System, ZILBS-10-6, Stent Diameter: 10mm, Stent Length: 6cm, Introduction System: 7 FR., Disposable-Single Use Only, Rx Only, Cook GI Endoscopy, Winston-Salem, NC. The device is used in palliation neoplasms in the biliary tree.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
A section of the introduction system may detach after the stent has deployed.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The consignee was notified by letter on/about 09/11/2008. The distributor was instructed to contact their customers and coordinate the return of all affected product to Cook Endoscopy. A responds form was attached to the letter to be completed and returned to Cook Medical to confirm receipt of notification and indicate quantity to be returned. Contact Cook Endoscopy at 1-336-744-0157, ext. 6325 for assistance.
Quantity in commerce
12
Distribution
USA: IL and OUS: China
Product codes and lots
Lot number W2528796