Z-1517-2008

GE Healthcare Innova 4100 Digital Fluoroscopic Imaging System with Bolus Chasing option

GE Healthcare / initiated 2007-10-21 / Terminated / root cause: manufacturing process

GE Healthcare began a recall of GE Healthcare Innova 4100 Digital Fluoroscopic Imaging System with Bolus Chasing option on . The reason FDA records is "Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1517-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare Innova 4100 Digital Fluoroscopic Imaging System with Bolus Chasing option.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option) may be underestimated by up to 50%. In order to avoid this issue, it is necessary to perform a system reset or reboot every day. GE Healthcare is working on a c
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
GE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed at their site at no cost.
Quantity in commerce
146 total
Distribution
Worldwide Distribution: USA, Argentina, Australia, Canada, China, Germany, Japan, Mexico and Russia.
Product codes and lots
8613100, 77573841, 0910064009, 3018774100, 4103323100, 4103374100, 4103623100, 4105434100, 4438494100, 5138534100, 5703214100, 7083333100, 7278253100, 8019643100, 8595724100, 9408984100, 082416040017, 082416070004, 082416120006, 082416130013, 082416140003, 201541LAB2, 229228INNOV41V, 239436CATH1, 251435ICV5, 270688INN, 304255INNOVA, 304347INNOVA, 309691P3100, 310423INNOVA1, 313593IC1, 313593IC2, 314525SACL2, 315448RM4, 315448RM5, 316268INNOVAA, 319338VA4100, 325670CATH, 334286BSP, 352333VASC, 352376SP, 352401CATH, 352401VASC, 378INV31, 407SPECIALS, 409899LAB3, 413447CATH, 414961INN4100, 416756INNOVA, 417269SCLD, 478633INOV, 502587LAB4, 508856OR41, 516562INNOVA4, 516663INNOVA2, 518525LAB1, 540953INNOVA, 561266IN31, 563421CATHLAB8, 573632CRA, 573761SMA, 573815BCR1, 574535IN41, 574647IN41, 6012884100SP1, 601553CL3, 602239LAB4, 602239LAB6, 603778ECL1, 606833INNOVA, 614293LIBERTY1, 614293LIBERTY2, 615338CV2, 615769SKYCV2, 619340VA8, 619482INNOVA1, 619541INNOVA31, 619552INNOVA1, 626359INNOVA, 702616INNOVA, 7028803100A, 706774CCL5, 713794LAB7, 713794LAB9, 715369INN4100, 717217INN31, 717763INNOVA41, 717763OR4100, 718283INNOVA3, 718518CVL1, 724656ANG, 724773C1, 727725SPEC1, 731422CV1, 732294INOV, 757398INNOVA, 757736INNOVA, 765453INNOVA1, 770991CL3, 773792CL3100, 773NORWG3100, 803329CL31, 803329CL41, 804594CATH, 808547QCV2A, 812238LIB1, 812279INNOVA1, 812283IN31, 812450EP1, 812482INNOVA, 813972INTV, 816276SP1, 817255NH4100, 817433HSW3100, 83INNOVA1188379, 845365V19, 850678IN3100, 865541CCL5, 865693CCL3, 901726SP2, 903CSM4100, 904202CATH3, 904276SP, 908522LAB1, 909788INNOVA1, 918307SHIN4, 918744JC4100, 928779INNOVA1, 936266CL1, 956389INN31, 956632INN4100, 956994INN1, 956994INN2, 972487ANGIO1, 973971LAB3, BPINNOVA2, BPINNOVA3R, BPINNOVA3R, ECXV1406C, FHORLCL1, HC1536XR04, RU1550VA01, RU2457VA01, SLIN4100, YV0022, and YV0036.