Z-0368-2008
GE Healthcare OEC 9900 mobile fluoroscopic x-ray system
GE Healthcare / initiated 2007-12-14 / Terminated / root cause: software
OEC Medical Systems, Inc began a recall of GE Healthcare OEC 9900 mobile fluoroscopic x-ray system on . The reason FDA records is "Possible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0368-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare OEC 9900 mobile fluoroscopic x-ray system
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- Possible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default selection to another selection. Manual technique modes are not affected.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE OEC sent a letter to customers December 15, 2007, and scheduled field representative visits for corrective software solution installation.
- Quantity in commerce
- 710
- Distribution
- Worldwide distribution: USA, Finland, Netherlands, Germany, Switzerland, France, United Kingdom, Australia, Italy, Ireland, Spain, Sweden, Belgium, New Zealand, China, Singapore, Poland, Norway, and India.
- Product codes and lots
- All codes