Z-0368-2008

GE Healthcare OEC 9900 mobile fluoroscopic x-ray system

GE Healthcare / initiated 2007-12-14 / Terminated / root cause: software

OEC Medical Systems, Inc began a recall of GE Healthcare OEC 9900 mobile fluoroscopic x-ray system on . The reason FDA records is "Possible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0368-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare OEC 9900 mobile fluoroscopic x-ray system
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
Possible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default selection to another selection. Manual technique modes are not affected.
Root cause tag
software
FDA root cause
Software design
Action
GE OEC sent a letter to customers December 15, 2007, and scheduled field representative visits for corrective software solution installation.
Quantity in commerce
710
Distribution
Worldwide distribution: USA, Finland, Netherlands, Germany, Switzerland, France, United Kingdom, Australia, Italy, Ireland, Spain, Sweden, Belgium, New Zealand, China, Singapore, Poland, Norway, and India.
Product codes and lots
All codes