GE Healthcare Innova 4100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System
Datex - Ohmeda, Inc began a recall of GE Healthcare Innova 4100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System on . The reason FDA records is "GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1522-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Innova 4100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- An Urgent Medical Device Correction letter, dated June 2008 was sent to customers The letter described the issue, affected products and provided short term user recommendations. GE developed a new software release that corrects this condition and will launch the corrective action to upgrade all affected systems.
- Quantity in commerce
- 87
- Distribution
- Worldwide Distribution - AL, AZ, CA, FL,, GA, HI, IA,IL, IN, KY, LA,MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV, Puerto Rico. OUS to include: Australia, Byelarus, China, Denmark, France, Germany, Greece, Honduras, Iran, Iraq, Italy, Japan, Latvia, Lebanon, Mexico, Poland, Russia, Saudi Arabia, Turkey, United Kingdom, and Venezuela.
- Product codes and lots
- Systems id: 77573841, 80158241, 910063039, 3018774100, 4103374100, 4438494100, 5134754100, 5138534100, 6029234100, 8159374100, 8595724100, 9408984100, 82406010011, 082416130011, 82416130012, 00095VAS07, 205759CL2, 216844INNOVA, 229228INNOV41V, 229434INNOV, 239436CATH1, 251435ICV5, 304255INNOVA, 304347INNOVA, 315448RM4, 315448RM5, 319338VA4100, 325670CATH, 337981INNOVA, 352333VASC, 352376SP, 352401CATH, 352401VASC, 407518SP2, 407518SP3, 407SPECIALS, 414961INN4100, 478633INOV, 508856OR41, 515263IL4100, 516562INNOVA4, 573632CRA, 573761SMA, 574535IN41, 600050VA01, 6012884100SP1, 606833INNOVA, 609652CATH3, 614293LIBERTY1, 614293LIBERTY2, 704487CR41, 706802FINN, 715369INN4100, 717763INNOVA41, 717763OR4100, 718518CVL1, 724656ANG, 727725SPEC1, 727819SPSC, 740592INNOVA, 757398INNOVA, 757736INNOVA, 803329CL41, 804320CATH2B, 804594CATH, 812279INNOVA1, 813972INTV, 816276SP1, 817255NH4100, 828696IN41, 865541CCL5, 865693CCL3, 901726SP2, 903CSM4100, 904276SP, 906485ANGIO, 918307SHIN4, 941483INNOVA41, 956632INN4100, A5630803, B5801711, BE5005VA01, BY4020VA01, RU1018VA03, SA1188VA03, SLIN4100, YV0003.