Z-0161-2009

Accessory Head Rest for Alphastar Operating Tables

Maquet Cardiovascular Us Sales, LLC / initiated 2008-09-26 / Terminated

MAQUET Inc. began a recall of Accessory Head Rest for Alphastar Operating Tables on . The reason FDA records is "During installation or cleaning there is a risk of crushing the finger by handling the head rest in the area of the gas strut while simultaneously actuating the release lever". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0161-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Accessory Head Rest for Alphastar Operating Tables
Product code
FQO Table, Operating-Room, Ac-Powered
Reason for recall
During installation or cleaning there is a risk of crushing the finger by handling the head rest in the area of the gas strut while simultaneously actuating the release lever.
FDA root cause
Other
Action
Urgent Device Correction letters were sent beginning September 26, 2008 via Fed Ex. Letters identified the potential for finger crush in conjunction with the head rest. Proper operating instructions are provided as well as a warning sticker to be applied to the head rest indicating proper positioning. All letters included a customer response form to be completed by each institution. Information requested is: that device correction letter has been read and instructions followed; that the corrected operating instructions have been received; that the warning sticker has been properly placed on the headrest, or that they do not have the head rest identified in the device correction letter. Questions should be addressed to USA Maquet Inc at 1-888-627-8383 #1.
Quantity in commerce
88 units in US; 1884 units worldwide
Distribution
Nationwide Distribution
Product codes and lots
USA Part Number 1130.67F0, No serial numbers