Z-1090-2009

BD PrepStain System, Model Number: 799-14000-00; Catalog Number: 05CR000021, 05CR00021R, 799-13000-00...

BD Diagnostic Systems Tripath / initiated 2008-09-26 / Terminated / root cause: software

Bd Diagnostic Systems Tripath began a recall of BD PrepStain System, Model Number: 799-14000-00; Catalog Number: 05CR000021, 05CR00021R, 799-13000-00, 799-13000-00R, 799-14000-00 and 799-14000-00R; Catalog... on . The reason FDA records is "Software Error in the non-gyn program of some devices may cause non-gyn samples to be incorrectly transferred from centrifuge tubes to slides potentially result in mismatched patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1090-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BD Diagnostic Systems Tripath
Product
BD PrepStain System, Model Number: 799-14000-00; Catalog Number: 05CR000021, 05CR00021R, 799-13000-00, 799-13000-00R, 799-14000-00 and 799-14000-00R; Catalog Number: 05CR000021, 05CR000021R, 799-13000-00, 799-13000-00R, 799-14000-00 and 7914000-00R. PrepStain is a liquid-based thin layer cell preparation device for preparation and staining of Prep smear slides from cervical specimens in a liquid based format (gyn use) and of cytology examination slides from non-cervical specimens such as urine and lung lavage.
Product code
MKQ Processor, Cervical Cytology Slide, Automated
Reason for recall
Software Error in the non-gyn program of some devices may cause non-gyn samples to be incorrectly transferred from centrifuge tubes to slides potentially result in mismatched patient results.
Root cause tag
software
FDA root cause
Software change control
Action
BD Diagnostics began contacting known non-gyn customers by phone on 09/26/2008. Customers were notified by Urgent - Important Product Information letter on 09/29/2008. They were advised to stop using the system until they complete a list of actions provided in the letter. A telephone contact number for Technical Support was provided in the letter. A permanent software based solution will be provided to customers when completed. They anticipate the software will be available in October 2008.
Quantity in commerce
662 units
Distribution
Worldwide Distribution
Product codes and lots
GYN software version 1.2, Non-GYN version 2.80; GYN software version 1.3.0.1, Non-GYN software version 2.80.