Z-0273-2009

Gyrus ACMI, Inc. dissector PlasmaKnife (DPK) Catalog No: 7035-3005 A single use only bipolar electrosurgical...

Gyrus ACMI, Inc. / initiated 2008-10-01 / Terminated / root cause: manufacturing process

Gyrus ACMI Corporation began a recall of Gyrus ACMI, Inc. dissector PlasmaKnife (DPK) Catalog No: 7035-3005 A single use only bipolar electrosurgical instrument with the ability to cut and coagulate... on . The reason FDA records is "Sterility may be compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0273-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gyrus ACMI, Inc.
Product
Gyrus ACMI, Inc. dissector PlasmaKnife (DPK) Catalog No: 7035-3005 A single use only bipolar electrosurgical instrument with the ability to cut and coagulate soft tissue in head and neck surgery within an ambient air environment.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Sterility may be compromised
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Gyrus ACMI initiated a voluntary recall by an Urgent: Medical Device Recall letter on 10/01/08 informing accounts of the problem, cease any further use of the affected product, remove any of the affected product service representative to obtain a Return Goods Authorization, return the product, and complete the reply form and fax it to the Regulatory Affairs Dept. so that the response can be tracked.
Quantity in commerce
1340 units
Distribution
Worldwide Distribution --- including USA and countries of Canada, UK, Italy, Spain, Norway, Australia, and Taiwan.
Product codes and lots
Lot Numbers: 0805064, 0805102, 0805103, 0806030, 0806039, 0806187, and 0807021