Z-0314-2009
AXIOM Artis System dFA using software version VB31D with 30x40 cm2 flat detectors. AXIOM Artis is a...
Siemens Medical Solutions USA, Inc. / initiated 2008-08-25 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of AXIOM Artis System dFA using software version VB31D with 30x40 cm2 flat detectors. AXIOM Artis is a angiography system developed for single and biplane... on . The reason FDA records is "System may switch to emergency fluoroscopy mode". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0314-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- AXIOM Artis System dFA using software version VB31D with 30x40 cm2 flat detectors. AXIOM Artis is a angiography system developed for single and biplane diagnostic imaging and interventional procedures.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- System may switch to emergency fluoroscopy mode.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- A Customer Safety Advisory Letter dated 8/20/08 was issued with Update Instructions AX045/08/S. The letter informed the customer of the potential issue and provided instruction to avoid its occurrence. A software update (VB31E) was released to correct the issue.
- Quantity in commerce
- 2 units
- Distribution
- Nationwide Distribution including states of AR, CA, CO, FL, IA, IL, IN, KS, KY, MI, NY, TX, VA, and WY.
- Product codes and lots
- Model number 7555373. Serial numbers: 50176 and 51707