Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0121-2009

Axiom Artis dBC-M, Axiom Artis Modular Angiography X-Ray System, Model number 5917054

Siemens Medical Solutions USA, Inc. / initiated 2008-09-02 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Axiom Artis dBC-M, Axiom Artis Modular Angiography X-Ray System, Model number 5917054 on . The reason FDA records is "Incorrect contour finding: Image may be calibrated to the wrong size catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0121-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Axiom Artis dBC-M, Axiom Artis Modular Angiography X-Ray System, Model number 5917054.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
Root cause tag
software
FDA root cause
Software design
Action
Siemens issued a Customer Safety Advisory Notice dated 8/29/08 to inform customers of the potential issue. The recalling firm issued Update Instructions AX040/08/S to the service engineers which provides information to fix the problem.
Quantity in commerce
3 units
Distribution
Nationwide.
Product codes and lots
Model number 5917054. Serial numbers: 49001, 49008, and 49009.