Z-0123-2009
Axiom Artis dTC, Axiom Artis Modular Angiography X-Ray System, Model number 7413078
Siemens Medical Solutions USA, Inc. / initiated 2008-09-02 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Axiom Artis dTC, Axiom Artis Modular Angiography X-Ray System, Model number 7413078 on . The reason FDA records is "Incorrect contour finding: Image may be calibrated to the wrong size catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0123-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Axiom Artis dTC, Axiom Artis Modular Angiography X-Ray System, Model number 7413078.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens issued a Customer Safety Advisory Notice dated 8/29/08 to inform customers of the potential issue. The recalling firm issued Update Instructions AX040/08/S to the service engineers which provides information to fix the problem.
- Quantity in commerce
- 49 units
- Distribution
- Nationwide.
- Product codes and lots
- Model number 7413078. Serial numbers: 46129, 46189, 46190, 46206, 46208, 46209, 46210, 46211, 46212, 46213, 46214, 46215, 46216, 46217, 46218, 46219, 46220, 46221, 46222, 46223, 46224, 46225, 46226, 46227, 46228, 46229, 46230, 46231, 46232, 46233, 46234, 46235, 46236, 46237, 46238, 46239, 46241, 46242, 46243, 46244, 46245, 46246, 46248, 46249, 46250, 46251, 46253, 46254, and 46255.