Z-0124-2009
Axiom Artis dBA, Axiom Artis Modular Angiography X-Ray System, Model number 7555357
Siemens Medical Solutions USA, Inc. / initiated 2008-09-02 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Axiom Artis dBA, Axiom Artis Modular Angiography X-Ray System, Model number 7555357 on . The reason FDA records is "Incorrect contour finding: Image may be calibrated to the wrong size catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0124-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Axiom Artis dBA, Axiom Artis Modular Angiography X-Ray System, Model number 7555357.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens issued a Customer Safety Advisory Notice dated 8/29/08 to inform customers of the potential issue. The recalling firm issued Update Instructions AX040/08/S to the service engineers which provides information to fix the problem.
- Quantity in commerce
- 93 units
- Distribution
- Nationwide.
- Product codes and lots
- Model number 7555357. Serial numbers: 53002, 53007, 53008, 53027, 53028, 53030, 53031, 53032, 53046, 53059, 53079, 53089, 53090, 53091, 53092, 53105, 53106, 53112, 53122, 53125, 53127, 53129, 53134, 53137, 53148, 53151, 53157, 53158, 53163, 53167, 53168, 53169, 53170, 53180, 53182, 53183, 53184, 53185, 53187, 53188, 53189, 53190, 53191, 53192, 53193, 53195, 53196, 53197, 53199, 53200, 53201, 53202, 53203, 53204, 53205, 53206, 53207, 53208, 53209, 53210, 53211, 53212, 53213, 53215, 53216, 53217, 53218, 53219, 53220, 53221, 53222, 53223, 53224, 53225, 53226, 53227, 53228, 53229, 53231, 53232, 53233, 53234, 53235, 53237, 53238, 53239, 53240, 53706, 53707, 53708, 53709, 53710, and 53711.