Z-0125-2009
Axiom Artis dFC-M, Axiom Artis Modular Angiography X-Ray System, Model number 7727717
Siemens Medical Solutions USA, Inc. / initiated 2008-09-02 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Axiom Artis dFC-M, Axiom Artis Modular Angiography X-Ray System, Model number 7727717 on . The reason FDA records is "Incorrect contour finding: Image may be calibrated to the wrong size catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0125-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Axiom Artis dFC-M, Axiom Artis Modular Angiography X-Ray System, Model number 7727717.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens issued a Customer Safety Advisory Notice dated 8/29/08 to inform customers of the potential issue. The recalling firm issued Update Instructions AX040/08/S to the service engineers which provides information to fix the problem.
- Quantity in commerce
- 11 units
- Distribution
- Nationwide.
- Product codes and lots
- Model number 7727717. Serial numbers: 40000, 40061, 40062, 40063, 40064, 40065, 40066, 40067, 40069, 40070, and 40071