Z-0126-2009
Axiom Artis dFA, Axiom Artis Modular Angiography X-Ray System, Model number 7555373
Siemens Medical Solutions USA, Inc. / initiated 2008-09-02 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Axiom Artis dFA, Axiom Artis Modular Angiography X-Ray System, Model number 7555373 on . The reason FDA records is "Incorrect contour finding: Image may be calibrated to the wrong size catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0126-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Axiom Artis dFA, Axiom Artis Modular Angiography X-Ray System, Model number 7555373.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens issued a Customer Safety Advisory Notice dated 8/29/08 to inform customers of the potential issue. The recalling firm issued Update Instructions AX040/08/S to the service engineers which provides information to fix the problem.
- Quantity in commerce
- 41 units
- Distribution
- Nationwide.
- Product codes and lots
- Model number 7555373. Serial numbers: 50080, 50135, 50137, 50138, 50139, 50140, 50141, 50142, 50143, 50144, 50145, 50146, 50147, 501 48, 50149, 50151, 50152, 50153, 50154, 50155, 50156, 50157, 50158, 50159, 50160, 50161, 50163, 50164, 50165, 50166, 50167, 50169, 50170, 50171, 50172, 51708, 51709, 51710, 51711, 51712, and 51713.