Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0427-2009

Dual Port Wizard Low Profile Replacement Gastrostomy Device, Code 00213W, 20F, 2.4cm, 20cc, Sterile EO, Bard...

Bard Access Systems, Inc. / initiated 2008-09-23 / Terminated / root cause: design

C R Bard Inc began a recall of Dual Port Wizard Low Profile Replacement Gastrostomy Device, Code 00213W, 20F, 2.4cm, 20cc, Sterile EO, Bard Access Systems, Inc., Salt Lake City, UT 84116 on . The reason FDA records is "Gastrostomy device anti-reflux valve may allow leakage from the stomach". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0427-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Access Systems, Inc.
Product
Dual Port Wizard Low Profile Replacement Gastrostomy Device, Code 00213W, 20F, 2.4cm, 20cc, Sterile EO, Bard Access Systems, Inc., Salt Lake City, UT 84116.
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
Gastrostomy device anti-reflux valve may allow leakage from the stomach.
Root cause tag
design
FDA root cause
Component design/selection
Action
Customers notified by letter (Product Field Action letter) on September 23, 2008 and instructed to remove and return any affected product to Bard Access Systems, Inc (BAS). Customers were instructed in the recall letter to respond by completing and faxing the enclosed Reply Form and Inventory Reconciliation Form. For assistance, contact BAS customer service at 800-290-1689.
Quantity in commerce
1655 units
Distribution
Worldwide Distribution including US, Latin America and countries of Canada, Japan, Finland, Norway, Sweden, Austria, Belgium, Holland, Great Britain, Italy, France, Switzerland, Spain, Germany.
Product codes and lots
43APA041,  43BQA048 43bQA187,  43DOA201,  43DPA050,  43DPA215,  43DQA067,  43EOA048,  43FOA117,  43FPA178,  43FQA042,  43GPA151,  43GPA212R,  43GPA236R,  43GQA035,  43HOA037,  43HPA056,  43HQA033,  43IOA046,  43IPA048,  43IQA032,  43JNA049,  43JNA222,  43JOA023,  43JPA043,  43KPA058,  43KQA037,  43LNA057,  43LOA034,  HURD0675,  HURD1929,  HURE2907,  HURF0500,  HURH1093,  HURJ0608,  HURJ1591, and  HURK1515.