Z-2461-2010

Nuclear Magnetic Resonance Imaging System (MRI). Trace item 9896-030-14351 -1.5T Sense XL Torso coil 12nc on...

Philips North America LLC / initiated 2008-09-30 / Terminated

Philips Medical Systems North America Co. Phillips began a recall of Nuclear Magnetic Resonance Imaging System (MRI). Trace item 9896-030-14351 -1.5T Sense XL Torso coil 12nc on coil: 4535-671-4188x on . The reason FDA records is "The cables of the coil may become too hot and cause burns when placed too close to the patient. While using the 1.5T Sense XL Torso coil, unwanted radiofrequencies (RF) interaction with the patient can occur. Especially...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2461-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Nuclear Magnetic Resonance Imaging System (MRI). Trace item 9896-030-14351 -1.5T Sense XL Torso coil 12nc on coil: 4535-671-4188x.
Product code
MOS Coil, Magnetic Resonance, Specialty
Reason for recall
The cables of the coil may become too hot and cause burns when placed too close to the patient. While using the 1.5T Sense XL Torso coil, unwanted radiofrequencies (RF) interaction with the patient can occur. Especially in combination with high SAR levels and when the RF cables of the coil are too close to the patient, this RF interaction has resulted in RF burns to patients.
FDA root cause
Other
Action
Philips Healthcare issued URGENT-Device Correction Notices dated September 30, 2008 to their consignees. The notice identified the recalled device as 1.5 Intera and 1.5T Achieva MRI systems, described the risk to patients, and the actions to be taken by customers.. The firm has issued Field Change Order (FCO) 78100242. Field service representatives will visit each consignee and conduct a hardware upgrade, applying a sleeve over the cable that will ensure proper spacing between the cable and patient, reducing the chance of RF interaction and burns. The consignees should contact the Philips Call Center at 1-800-772-9377, #5 #3, #1 and reference "FCO 78100242 - PATIENT CABLE SLEEVE FOR 1.5T SENSE XL TORSO COIL".
Quantity in commerce
47 units total (Achieva 1.5T and Intera 1.5T)
Distribution
Nationwide Distribution: Throughout USA. No distribution to foreign consignees.
Product codes and lots
Devices are identified as Site numbers: 544006, 537871, 534961, 544704, 545154, 545236, 545219, 545943, 546637, 545879, 545914, 543286, 547585, 548729, 550876, 553379, 550044, 554797, 41443789, 555412, 41444993, 41445232, 556803, 557348, 557344, 557513, 557179, 557644, 557560, 551656, 558845, 41446147, 552550, 552516, 555787, 41445636, 41455922, 83325, 558369, 86968, 101408, 101877, 102110, 102353, 102004, 104343, and 41455948.