Dimension EXL Clinical Chemistry System used with software versions 8.5.1 and 8.5.1SP3. The system is a...
Dade Behring, Inc. began a recall of Dimension EXL Clinical Chemistry System used with software versions 8.5.1 and 8.5.1SP3. The system is a discrete, random-access, microprocessor-controlled... on . The reason FDA records is "Error messages, "Wash Aspirator Failure" and "Wash station unable to recover" are not generated when one or both HM wash probes are totally occluded". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0581-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension EXL Clinical Chemistry System used with software versions 8.5.1 and 8.5.1SP3. The system is a discrete, random-access, microprocessor-controlled, integrated instrument/chemistry system that measures a variety of analytes, including enzyme activities, in body fluids for in vitro diagnostic use.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Error messages, "Wash Aspirator Failure" and "Wash station unable to recover" are not generated when one or both HM wash probes are totally occluded.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm notified consignees by telephone to inform them of the problem. A follow-up letter, "Urgent Field Safety Notice" dated 9/30/08 was issued with information describing the problem, the possible impact of an occluded wash probe on assay results, and emphasizing regular performance of routine wash probe maintenance as detailed in the operator guide to prevent wash probe occlusion. For questions, please contact the Siemens Technical Solution Center at 800-441-9250.
- Quantity in commerce
- 57 units
- Distribution
- Nationwide Distribution including states of AR, CA, CO, IA, ID, IL, IN, LA, MD, MI, MO, MS, NC, NE, OH, OK, PA, SD, TN, TX, VA, and WI.
- Product codes and lots
- Software versions 8.5.1 and 8.5.1SP3.