Z-0484-2009

STERRAD NX Sterilizer, Product Code 10033 Low temperature sterilizers that use hydrogen peroxide gas plasma...

Advanced Sterilization Products / initiated 2008-10-07 / Terminated / root cause: manufacturing process

Advanced Sterilization Products began a recall of STERRAD NX Sterilizer, Product Code 10033 Low temperature sterilizers that use hydrogen peroxide gas plasma to inactivate microorganisms on a broad range of... on . The reason FDA records is "ASP has discovered a component defect in some of the UV lamp power supplies used in certain STERRAD NX Sterilizers. This defect can potentially cause the Hydrogen Peroxide Monitor to give inaccurate readings". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0484-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Sterilization Products
Product
STERRAD NX Sterilizer, Product Code 10033 Low temperature sterilizers that use hydrogen peroxide gas plasma to inactivate microorganisms on a broad range of medical and surgical instruments.
Product code
MLR Sterilizer, Chemical
Reason for recall
ASP has discovered a component defect in some of the UV lamp power supplies used in certain STERRAD NX Sterilizers. This defect can potentially cause the Hydrogen Peroxide Monitor to give inaccurate readings.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The Urgent: Product Correction customer notifications, dated October 6, 2008, were hand delivered by field representatives starting October 7, 2008. The letter advises that Advanced Sterilization Products has discovered a component defect in some of the UV lamp power supplies used in certain STERRAD Sterilizers. This defect can potentially cause the Hydrogen Peroxide Monitor to give inaccurate readings. A list of the affected sterilizer serial numbers is included with the letter. The letter advises that the Field Service Engineer has come to their facility to inspect and, if necessary, replace this power supply. Customers are asked to sign the letter acknowledging receipt of this letter upon completion of the work by the Field Service Engineer.
Quantity in commerce
32 potentially affected systems in the US - 20 installed, 5 uninstalled, 7 potentially affected. There are also 15 units internationally.
Distribution
Worldwide Distribution --- USA including states of WV, CA, TN, AZ, WI, NC, MO, NJ, DE, GA, SC, WA, FL, MA, NM, MI, and AL and countries of Ecuador, China, Israel, Japan, Korea, Middle East, Australia, Czech Republic, Spain, Brazil, and Russia.
Product codes and lots
Serial Numbers: 0033081006, 0033081007, 0033081016, 0033081017, 0033081019, 0033081020, 0033081022- 0033081029, 0033081036, 0033081039, 0033081048, 0033081059, 0033081080, 0033081082, 0033081086, 0033081088, 0033081089, 0033081091, 0033081095, 10033060892, 10033060379, 100330050671, 10033070576,10033080067,10033050663, and 10033050661.