Z-1447-2012

ARCHITECT/AEROSET Activated Aspartate Aminotransferase The Activated Asparate Aminotransferase (Activated...

Abbott Medical / initiated 2008-06-30 / Terminated / root cause: manufacturing process

Abbott Laboratories Inc. began a recall of ARCHITECT/AEROSET Activated Aspartate Aminotransferase The Activated Asparate Aminotransferase (Activated AST) assay is used for the quantitation of asparate... on . The reason FDA records is "This letter is to inform you that some samples may generate A-AST results of less than (<) 5 U/L or results within the reference range when actual concentration exceeds the non-Flex linearity of 1985 U/L. In other cases...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1447-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT/AEROSET Activated Aspartate Aminotransferase The Activated Asparate Aminotransferase (Activated AST) assay is used for the quantitation of asparate aminotransferse in human serum or plasma
Product code
CIT Nadh Oxidation/Nad Reduction, Ast/Sgot
Reason for recall
This letter is to inform you that some samples may generate A-AST results of less than (<) 5 U/L or results within the reference range when actual concentration exceeds the non-Flex linearity of 1985 U/L. In other cases the following errors may be generated. -ARCHITECT cSystems: Error Code 1053 (Unable to calculate result, rate reaction linearity failure) -AEROSET: RL% Error Code Revised P
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Abbott Laboratories sent a Product Correction letter dated August 5, 2008, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to do the following: Update your assay parameters for A-AST 1. Follow the steps in the appropriate attachment: - Use Attachment A for ARCHITECT cSystems - Use Attachment B for AEROSET. 2. All A-AST results of less than (<) 5 U/L should be evaluated by comparing to other clinically relevant assays Retain a copy of this letter for your laboratory records. For questions regarding this recall call 626-440-0705.
Quantity in commerce
1,646
Distribution
Worldwide Distribution including Germany
Product codes and lots
List Numbers 8L91-20: 57028HW00, 59035HW00, 61031HW00, 64052HW00; List Numbers 8L91-40: 57017HW00, 59031HW00, 61038HW00, 64050HW00