K-ASSAY Alpha-1 Microglobulin Reagent Kit. Kit contains: 1 vial Reagent 1, Buffer 50 mL, and 1 vial reagent...
Kamiya Biomedical Company, LLC began a recall of K-ASSAY Alpha-1 Microglobulin Reagent Kit. Kit contains: 1 vial Reagent 1, Buffer 50 mL, and 1 vial reagent 2, Antibody Reagent 50 mL. The product is packaged... on . The reason FDA records is "Standardization of the reagent performance data did not match the standardization used to assign the calibrator values". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0923-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Kamiya Biomedical Company, LLC
- Product
- K-ASSAY Alpha-1 Microglobulin Reagent Kit. Kit contains: 1 vial Reagent 1, Buffer 50 mL, and 1 vial reagent 2, Antibody Reagent 50 mL. The product is packaged in a kit box along with with the package insert. The package insert is labeled in part: "K-ASSAY Alpha-1 Microglobulin...Cat. No. KAI-056...For the Quantitative Determination of human Alpha-1 Microglobulin in urine, serum, and plasma...KAMIYA BIOMEDICAL COMPANY...12779 Gateway Drive...Seattle, WA 98168 USA...".
- Product code
- MGA Alpha-1 Microglobulin, Antigen, Antiserum, Control
- Reason for recall
- Standardization of the reagent performance data did not match the standardization used to assign the calibrator values.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Kamiya Biomedical Company, LLC sent an urgent medical Device Recall letter dated September 17, 2008, and the corrected package insert via e-mail. . The letter identified the product, the problem, and the action the customer should take. Customers were instructed to replace their previous package inserts with the corrected version and asked to sign and FAX the letter back once that had been completed. As one customer is also a distributor, the recalling firm told the this distributor's representative to advise their single customer of this correction and replace the package insert, and return the signed letter back to the firm. On September 22, 2008, both customers were contacted by phone and reminded to fax the signed letters once completed. For questions regarding this recall call 1-800-222-0342.
- Quantity in commerce
- 5 kits
- Distribution
- Kits were distributed to one laboratory in MN and to one distributor in PA who further distributed product to one customer.
- Product codes and lots
- Lot number 418041, catalog number KAI-056.