Z-0322-2009

Arjo Maxi 500 Patient Lift with 2 Point Spreader Bar and Scale; a Non-AC-Powered Patient Lift; Model...

Arjo, Inc. / initiated 2008-10-20 / Terminated / root cause: manufacturing process

Arjo, Inc. began a recall of Arjo Maxi 500 Patient Lift with 2 Point Spreader Bar and Scale; a Non-AC-Powered Patient Lift; Model KM560101; BHM Medical Inc., Magog, QC, Canada The Maxi 500... on . The reason FDA records is "The pivot bolt that attaches the hanger bar to the scale of the patient lift can break, resulting in the hanger bar falling". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0322-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arjo, Inc.
Product
Arjo Maxi 500 Patient Lift with 2 Point Spreader Bar and Scale; a Non-AC-Powered Patient Lift; Model KM560101; BHM Medical Inc., Magog, QC, Canada The Maxi 500 is a mobile passive lift intended to be for lifting and transferring of patients in hospitals, nursing homes, or other health care facilities.
Product code
FSA Lift, Patient, Non-Ac-Powered
Reason for recall
The pivot bolt that attaches the hanger bar to the scale of the patient lift can break, resulting in the hanger bar falling.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Arjo sent an Urgent Device Recall Customer Notification letter dated 10/20/08 to all affected end users to the attention of the Administrator/Risk Manager via UPS 2nd day on the 10/20/08. The letters informed the accounts of the potential for the hanger bar to detach from the scale due to shearing of the pivot bolt. The accounts were advised to review the enclosed Safety Advisory Notice, note the affected model and serial numbers, and avoid using the lift under load on a high friction floor like carpet or soft tile while using the hanger bar to move the lift. The accounts were advised that an Arjo Field Service Technician would contact them within two weeks to schedule a visit to the account to perform repairs on their Maxi 500 lifts. Any questions were directed to Arjo Quality Department at 800-323-1245, ext. 6118. The accounts were also instructed to complete and fax back to Arjo the enclosed Customer Response Form, providing the facility name and address, contact name and phone number, the numbers of affected units, and acknowledgment of receipt and understanding of the Urgent Device Correction.
Quantity in commerce
38 units
Distribution
Nationwide Distribution.
Product codes and lots
serial numbers KM5610713, KM5610729, KM5610737, KM5610114, KM5610115, KM5610550, KM5610711, KM5610551, KM5610693, KM5610694, KM5610696, KM5611003, KM5611004, KM5611005, KM5611006, KM5611007, KM5611008, KM5611009, KM5611010, KM5611011, KM5611012, KM5611013, KM5611014, KM5611015, KM5611016, KM5611017, KM5611018, KM5611019, KM5611020, KM5611021, KM5611022, KM5611023, KM5611024, KM5611025, KM5611026, KM5611027, KM5610813, and KM5610667.