Stryker Rotary Handpiece, Single Trigger, For use with Stryker System 6. Stryker Instruments, Kalamazoo, MI...

Stryker Corp. / initiated 2008-10-10 / Terminated / root cause: design

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker Rotary Handpiece, Single Trigger, For use with Stryker System 6. Stryker Instruments, Kalamazoo, MI; REF 6203-000-000S1. The device is intended for use... on . The reason FDA records is "The handpieces may continue to run even when the trigger is not depressed, which could result in the unintended cutting of bone or soft tissue damage". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1023-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker Rotary Handpiece, Single Trigger, For use with Stryker System 6. Stryker Instruments, Kalamazoo, MI; REF 6203-000-000S1. The device is intended for use in surgical procedures involving drilling, reaming, driving wire or pins, and cutting bone and hard tissue.
Product code
KIJ Instrument, Surgical, Orthopedic, Dc-Powered Motor And Accessory/Attachment
Reason for recall
The handpieces may continue to run even when the trigger is not depressed, which could result in the unintended cutting of bone or soft tissue damage.
Root cause tag
design
FDA root cause
Component design/selection
Action
Consignees were notified of the problem via letter dated October 10, 2008, provided interim instructions to reduce the risk, and informed that Stryker instruments will replace the trigger on all handpieces. Consignees were also instructed on the following: 1) Review the letter and distribute to all affected departments, 2) Complete and sign the enclosed Business Reply Form and fax to 336-499-8603 and 3) Call Stryker Instruments Service Department at 888-308-1983 to schedule service for the handpieces.
Quantity in commerce
25,312 of all units
Distribution
Worldwide Distribution including USA and countries of Australia, Brazil, Canada, Chile, Colombia, Costa Rica, England, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Romania, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan and Turkey.
Product codes and lots
Serial numbers 0817003173 through 0825905983.