Stryker Precision Oscillating Tip Saw, For use with Stryker System 6. Stryker Instruments, Kalamazoo, MI; REF...
Stryker Instruments Div. of Stryker Corporation began a recall of Stryker Precision Oscillating Tip Saw, For use with Stryker System 6. Stryker Instruments, Kalamazoo, MI; REF 6209. The product is an oscillating cutting... on . The reason FDA records is "The handpieces may continue to run even when the trigger is not depressed, which could result in the unintended cutting of bone or soft tissue damage". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1030-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker Precision Oscillating Tip Saw, For use with Stryker System 6. Stryker Instruments, Kalamazoo, MI; REF 6209. The product is an oscillating cutting device used for cutting bone and bone related tissue.
- Product code
- KIJ Instrument, Surgical, Orthopedic, Dc-Powered Motor And Accessory/Attachment
- Reason for recall
- The handpieces may continue to run even when the trigger is not depressed, which could result in the unintended cutting of bone or soft tissue damage.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Consignees were notified of the problem via letter dated October 10, 2008, provided interim instructions to reduce the risk, and informed that Stryker instruments will replace the trigger on all handpieces. Consignees were also instructed on the following: 1) Review the letter and distribute to all affected departments, 2) Complete and sign the enclosed Business Reply Form and fax to 336-499-8603 and 3) Call Stryker Instruments Service Department at 888-308-1983 to schedule service for the handpieces.
- Quantity in commerce
- 25,312 of all units
- Distribution
- Worldwide Distribution including USA and countries of Australia, Brazil, Canada, Chile, Colombia, Costa Rica, England, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Romania, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan and Turkey.
- Product codes and lots
- Serial numbers 0620602123 through 0825317713.