Z-0486-2009
FHC microTargeting Platform DBS Measuring Fixture, a component of the micro-targeting Drive System (Catalog...
FHC, Inc. / initiated 2008-10-08 / Terminated / root cause: manufacturing process
FHC, Inc. began a recall of FHC microTargeting Platform DBS Measuring Fixture, a component of the micro-targeting Drive System (Catalog Number 66-FA-SF). The device is a stereotactic... on . The reason FDA records is "Measuring fixture is incorrectly graduated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0486-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FHC, Inc.
- Product
- FHC microTargeting Platform DBS Measuring Fixture, a component of the micro-targeting Drive System (Catalog Number 66-FA-SF). The device is a stereotactic instrument used for the placement of recording and stimulating electrodes in the brain.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Measuring fixture is incorrectly graduated.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- FHC provided letter notification to FHC Field Representatives on 10/08/08 to hand deliver to customers. FHC Field Representatives will remove incorrectly marked product and provide replacement product. Direct questions to an FHC representative or FHC Quality and Regulatory Affairs at 207-666-5651, ext. 2210.
- Quantity in commerce
- 5 units
- Distribution
- Nationwide Distribution including states of CO, FL, NC and NM.
- Product codes and lots
- Not coded.