Fisher & Paykel Healthcare RT240 Adult Breathing Circuit Kit. lntended Use: The dual heated adult respiratory...
Fisher & Paykel Healthcare Inc began a recall of Fisher & Paykel Healthcare RT240 Adult Breathing Circuit Kit. lntended Use: The dual heated adult respiratory ventilator circuits are intended as conduits of... on . The reason FDA records is "The recall was initiated because certain lot dates of the RT240 Adult Breathing Circuit Kits manufactured on or before June 27, 2008 include a heated breathing circuit that may be more susceptible to damage when used in...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0414-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fisher & Paykel Healthcare, Inc.
- Product
- Fisher & Paykel Healthcare RT240 Adult Breathing Circuit Kit. lntended Use: The dual heated adult respiratory ventilator circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of the humidified gas, to reduce condensation.
- Product code
- BTT Humidifier, Respiratory Gas, (Direct Patient Interface)
- Reason for recall
- The recall was initiated because certain lot dates of the RT240 Adult Breathing Circuit Kits manufactured on or before June 27, 2008 include a heated breathing circuit that may be more susceptible to damage when used in excess of the specified seven-day maximum duration of use, which may increase the risk of malfunction or fire.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- Fisher & Paykel Healthcare, Inc. (FPH) issued an URGENT-PRODUCT RECALL letter (dated October 8, 2008) including a Customer recall letter for consignees/customers to Cardinal Health. Cardinal Health will, in turn, send the Customer recall letter dated October 10, 2008 to its consignees/customers. Fisher & Paykel Healthcare (FPH) also issued a similar set of recall notification letters to Norco on October 15, 2008. Recipients of the recall letter must perform the following: 1) Check all Cardinal Health stock of the product for the affected lots, 2) Immediately quarantine the affected product, 3) Complete a Device Recall Response Sheet, and 4) Fax or e-mail the completed Device Recall Response Sheet. Fisher & Paykel Healthcare, Inc.'s "Dear Customer Letter" for Cardinal Health and Norco instructs consignees to complete a customer response (i.e. fax enclosed Customer Acknowledgement Form) to FPH. The customer letter also provides instructions for product return. Fisher & Paykel Healthcare, Inc. recall communication instructions to Cardinal Health and Norco was for all affected product to be returned and quarantined by the distributors. The distributors will return any affected stock to Fisher & Paykel Healthcare, Inc. for documented destruction.
- Quantity in commerce
- 229,320 units
- Distribution
- Worldwide Distribution including US and country of Canada.
- Product codes and lots
- Lot Number: All lot numbers up to and including Lot 080627.