Boston Scientific Easy Core" Biopsy System, 15 ga. x 10 cm, REF 43-450, UPN M001434501, Sterilized with...
Boston Scientific Corporation began a recall of Boston Scientific Easy Core" Biopsy System, 15 ga. x 10 cm, REF 43-450, UPN M001434501, Sterilized with ethylene oxide gas . UPN for inner pouch M001434500... on . The reason FDA records is "Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0329-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Easy Core" Biopsy System, 15 ga. x 10 cm, REF 43-450, UPN M001434501, Sterilized with ethylene oxide gas . UPN for inner pouch M001434500 Indicated for use endoscopically or percutaneously to retrieve tissue sampling of soft organs, tumors or masses for histological analysis. Soft tissue sampling includes, but is not limited to, organs such as breast, liver, kidney or prostate.
- Product code
- FCG Biopsy Needle
- Reason for recall
- Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were sent a "Boston Scientific Urgent Medical Device Recall" letter dated October 21, 2008. The letter was addressed to The Hospital Field Action Contact. The letter described the problem and products being recalled. The letter stated that it was very important that the attached product table be used carefully and consider both inner and outer packaging product codes when searching for affected/recalled product as the UPN numbers on the inner and outer labeling are different. Requested to discontinue use of product, identify, and segregate and return affected products to BSC. Letter also requested to complete and return the Reply Verification Tracking Form.
- Quantity in commerce
- 3808 total for all products
- Distribution
- Nationwide Distribution --- including states of AL, AZ, AR, CA, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO,MT, NE , NV, NH, NH, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and PR.
- Product codes and lots
- 11524375 and 11535831