Z-0325-2009

FS-OMNI, Fusion Omni-Tome, Sterile, Rx only, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road...

Wilson-Cook Medical Inc. / initiated 2008-10-30 / Terminated / root cause: manufacturing process

Cook Endoscopy began a recall of FS-OMNI, Fusion Omni-Tome, Sterile, Rx only, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105, EC Representative: Cook... on . The reason FDA records is "The wire control port may have been omitted during manufacture". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0325-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Wilson-Cook Medical Inc.
Product
FS-OMNI, Fusion Omni-Tome, Sterile, Rx only, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105, EC Representative: Cook Ireland Ltd, O'Haioran Road, National Technology Park, Limerick, Ireland. The Fusion Omni-Tome is supplied sterile and is a single use device. This device is used in cannulation of the ductal system and for sphincterotomy.
Product code
KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Reason for recall
The wire control port may have been omitted during manufacture.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Cook Ireland (distributor) was contacted via email on 10/30/08 and was instructed to contact their customers and coordinate the return of all recall product to Cook Endoscopy. A response sheet was attached. Recall status reports will be issued monthly. All devices returned as part of this recall initiative will be dispositioned to scrap and destroyed. Should you have any questions, please contact Cook Endoscopy Customer Quality Assurance Department at 1-336-744-0157.
Quantity in commerce
11
Distribution
Ireland
Product codes and lots
Lot number W2501093