Vertical drive brake (on patient supports). Used on certain Philips Medical Systems MX series and Brilliance...
Philips Medical Systems (Cleveland) Inc began a recall of Vertical drive brake (on patient supports). Used on certain Philips Medical Systems MX series and Brilliance series CT scanners. The device is a vertical drive... on . The reason FDA records is "The product is being recalled because the patient support may travel downwards without being commanded to move". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0597-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Vertical drive brake (on patient supports). Used on certain Philips Medical Systems MX series and Brilliance series CT scanners. The device is a vertical drive brake that controls the vertical position of a patient support system or couch. The patient support system is intended to be used to place the patient in the correct horizontal and vertical position in order to be scanned. The couch is an accessory product to a computed tomography scanner.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- The product is being recalled because the patient support may travel downwards without being commanded to move.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A On 11/15/2007 & 6/6/2008 the firm sent Urgent Device Notification letters to their customers. They notified their customers that a Field Service Engineer will inspect and repair the brake system.
- Quantity in commerce
- 559
- Distribution
- Worldwide Distribution: USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and WV and countries of Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Canada, China, Colombia, Costa Rica, Czech Republic, Denmark, Egypt, England, France, Germany, Greece, China, India, Indonesia, Ireland, Iraq, Italy, Japan, Jordan, Korea, Lebanon, Libya, Lithuania, Malaysia, Malta, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, South Africa, South Korea, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Product codes and lots
- Assembly number 4535 664 98912.