Z-1091-2009
Glucose Oxidase Reagent as follows: Pointe Scientific Glucose Oxidase Reagent Set, Pointe Scientific, Inc...
Horiba Instruments Incorporated / initiated 2008-11-03 / Terminated / root cause: manufacturing process
Pointe Scientific, Inc. began a recall of Glucose Oxidase Reagent as follows: Pointe Scientific Glucose Oxidase Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog # G7519-1L, G7519-500 and... on . The reason FDA records is "The product is unable to maintain the specification for linearity through the shelf life of the product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1091-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- Glucose Oxidase Reagent as follows: Pointe Scientific Glucose Oxidase Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog # G7519-1L, G7519-500 and G7519-SAM. 2MM Diagnostics Glucose Oxidase Reagent Set, 2MM Corporation, Miami, FL; Reorder # RT-160-500. Pointe Scientific Glucose Oxidase Reagent, Pointe Scientific, Inc., Canton, MI; Catalog, MI; Catalog # 2-G7519-L, 7-G7519-100, 7-G7519-200, 7-G7519-1000, 8-G7519-100, and 12-RT-160-500. For the quantitative determination of Glucose in human serum. For in vitro diagnostic use only.
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- The product is unable to maintain the specification for linearity through the shelf life of the product.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified via an Urgent: Medical Device Recall letter dated 11/3/08 that all test results obtained with these lots are questionable and that attending physicians should be informed to determine whether additional testing is required. Also, consignees should examine their inventories for any of the lot numbers listed and dispose of them. If the product has been further distributed, identify customers and notify them of the product recall by including a copy of the recall letter. The Recall Return Response Form should be completed and returned and upon receipt Pointe Scientific will replace any disposed of reagent. Questions should be directed to the Technical Support Department at 1-800-757-5313.
- Quantity in commerce
- 1500 liters
- Distribution
- Worldwide Distribution -- USA, Hong Kong, Indonesia, Israel, Jordan, Mexico, Nigeria, Poland, Russia, Serbia, Sri Lanka, Trinidad, and Venezuela.
- Product codes and lots
- All lots beginning in 722601, Exp. 08/09. (Lots 722601, 722601-009, 722601-017, 722601-035, 722601-057, 722601-072, 722601-081, 722601-084, 722601-085, 722601-092, 722601-093, 722601-094, 722601-134, 722601-136, 722601-168, 722601-185, 722601-232, 722601-233, 722601-254, 722601-255, 722601-318, 722601-337, and 722601-361.)