Z-0934-2009

Posey Connecting Straps & Belts/Locking Cuffs. The device is used to limit the wearer's movement by securing...

J. T. Posey Co. / initiated 2008-09-08 / Terminated

J T Posey Company began a recall of Posey Connecting Straps & Belts/Locking Cuffs. The device is used to limit the wearer's movement by securing cuffs to a connecting strap or belt on . The reason FDA records is "The recall was initiated after an investigation conducted by the Posey Company determined that a patient may be able to catch the belt on a stationary object (wall, pipe, etc.) and use their body force to tear or break...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0934-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
J. T. Posey Co.
Product
Posey Connecting Straps & Belts/Locking Cuffs. The device is used to limit the wearer's movement by securing cuffs to a connecting strap or belt.
Product code
FMQ Restraint, Protective
Reason for recall
The recall was initiated after an investigation conducted by the Posey Company determined that a patient may be able to catch the belt on a stationary object (wall, pipe, etc.) and use their body force to tear or break the strap, creating freedom of arm movement. In some patient populations, such freedom of movement could pose an unwanted hazard of serious injury to a patient or caregiver.
FDA root cause
Other
Action
An "Urgent Medical Device Recall (Correction)" letter dated August 15, 2008 was mailed via US Postal Service to affected customers. The letter described the issue, instructed customers to secure a replacement label with the correct application technique over the picture in the 2008 Posey Catalog and provided a corrected Posey Locking Cuffs Instruction Sheet (19273). The Posey Company also provided updated literature for the Posey Locking and Non-Locking Connecting Straps 19270 and 19254. The Posey Company asked customers to complete and return the "Reply Fax-Back Acknowledgment" Form via fax (626-443-5418) or e-mail ([email protected]) within 7 days. Contact your Sales Representative with any questions or concerns.
Quantity in commerce
51,200 (Approximately)
Distribution
Worldwide Distribution including USA and countries of Australia, Belgium, Canada, France, Germany, Ireland, Italy, Korea, Lithuania and Netherlands.
Product codes and lots
Catalog Numbers: 2320, 2330, 2346, 2350, 2351, 2355, 2360, 2208, 2209, 2210, 2211, 2212 and 2213.