Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0410-2009

PALINDROME EMERALD Kit - Cuffed Dual Lumen Catheter with Pre-curved Shaft and Heparin Coating, REF...

Medtronic, Inc. / initiated 2008-09-19 / Terminated

Covidien began a recall of PALINDROME EMERALD Kit - Cuffed Dual Lumen Catheter with Pre-curved Shaft and Heparin Coating, REF 8888145069, 14.5 Fr/CH (4.85mm) x 23cm. TYCO HEALTHCARE... on . The reason FDA records is "Incorrect size of the Venetrac stylets (an optional accessory). They are too short and cannot be used". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0410-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
PALINDROME EMERALD Kit - Cuffed Dual Lumen Catheter with Pre-curved Shaft and Heparin Coating, REF 8888145069, 14.5 Fr/CH (4.85mm) x 23cm. TYCO HEALTHCARE GROUP LP, Mansfield, MA 02048. --- NOTE: Both the "sport pack" and "kit" package configurations include the catheter, Venetrac insertion stylets, pull apart sheath, and other accessories necessary for placement. The kit includes additional components that would also be available at the hospital, such as a scalpel, wound dressing, etc. The Venetrac insertion stylet is an optional accessory. The 510k for the Venetrac accessory is K051584. The 510k for the catheter is K060509.
Product code
NYU Catheter, Hemodialysis, Implanted, Coated
Reason for recall
Incorrect size of the Venetrac stylets (an optional accessory). They are too short and cannot be used.
FDA root cause
Mixed-up of materials/components
Action
Starting on 9/19/08, Covidien Sales Representatives visited each of the hospitals that received the catheters, informed the staff that the Venetrac insertion stylets were too short, and placed pink corrective sticker labels on any unused product. Should you require additional information, please contact [email protected] or 508-216-8532.
Quantity in commerce
82 units
Distribution
Nationwide distribution including states of OH, WI, LA, TX, UT, CA, OR, and WA.
Product codes and lots
Product code/REF Number 8888145069; Lot No: 809404 and 811763.