Measles IgG Enzyme Immunoassay Test Kit, DIAMEDIX. The assay is intended for the semi-quantitation of human...
Diamedix Corporation began a recall of Measles IgG Enzyme Immunoassay Test Kit, DIAMEDIX. The assay is intended for the semi-quantitation of human IgG antibodies to measles virus in human serum by... on . The reason FDA records is "Possible false-negative test results: The Sample A Diluent Lot # 50768 assigned to Is-Measles IgG Lot #'s 41108 and 41108B has higher conductivity than normal, which can sporadically interfere with the MAGO Plus/Aptus...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0607-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diamedix Corp.
- Product
- Measles IgG Enzyme Immunoassay Test Kit, DIAMEDIX. The assay is intended for the semi-quantitation of human IgG antibodies to measles virus in human serum by indirect immunoassay to aid in the assessment of the patient's immunological response to measles virus, to determine the immune status of individuals and, when evaluating paired sera, as an aid in the diagnosis of measles infection.
- Product code
- LJB Enzyme Linked Immunoabsorbent Assay, Rubeola Igg
- Reason for recall
- Possible false-negative test results: The Sample A Diluent Lot # 50768 assigned to Is-Measles IgG Lot #'s 41108 and 41108B has higher conductivity than normal, which can sporadically interfere with the MAGO Plus/Aptus sensing of and picking up of samples. As a result, if the sample is not picked up/processed, it could lead to the possibility of obtaining false negative results in some cases.
- FDA root cause
- Other
- Action
- Notification of all domestic customers via Certified Mail. Notification of the one international customer via Federal Express.
- Quantity in commerce
- 974Test Kits
- Distribution
- Worldwide Distribution --- including USA and country of England.
- Product codes and lots
- Lot number 41108 and 41108B