Z-0608-2009

Hitachi CXR4 Computed Tomography Scanner. This product is an x-ray imaging device that produces...

FUJIFILM Healthcare Americas Corporation / initiated 2008-11-05 / Terminated / root cause: software

Hitachi Medical Systems America Inc began a recall of Hitachi CXR4 Computed Tomography Scanner. This product is an x-ray imaging device that produces cross-sectional images of he body at different angles. The... on . The reason FDA records is "Software error: The device has a software error that can result in a condition where the displayed image orientation does not match the display orientation markers (right-left reversal)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0608-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Hitachi CXR4 Computed Tomography Scanner. This product is an x-ray imaging device that produces cross-sectional images of he body at different angles. The system reconstructs, processes, displays, and stores the collected images. The device output can provide an aid to diagnosis when used by a qualified physician and is intended for general purpose CT applications.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Software error: The device has a software error that can result in a condition where the displayed image orientation does not match the display orientation markers (right-left reversal).
Root cause tag
software
FDA root cause
Software design
Action
On 11/5/2008 Hitachi Medical Systems sent recall notification letters entitled "URGENT - DEVICE RECALL CORRECTION" to its customers informing them of the problem with the software. They stated that a Hitachi service representative will contact each customer to schedule a software upgrade. For further information, please contact Hitachi Medical Systems America Inc. by telephone at 330-425-1313.
Quantity in commerce
42
Distribution
Nationwide Distribution --- including states of AZ, CA, DE, FL, ID, IL, IN, KS, KY, LA, MA, MO, NJ, NY, OH, OK, PA, PR, SC, TX, VA, and WI.
Product codes and lots
Serial numbers: CXR46202 through CXR46248 excluding CXR46207, CXR46208, CXR46215, and CXR46222.