Roche/Hitachi Modular Analytics System, clinical chemistry analyzer, Modular P800 Module, Roche Diagnostics...
Roche Diagnostics Corp. began a recall of Roche/Hitachi Modular Analytics System, clinical chemistry analyzer, Modular P800 Module, Roche Diagnostics Corp., Indpls., IN; GMMI No. 04998669001. Is a... on . The reason FDA records is "If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1066-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche/Hitachi Modular Analytics System, clinical chemistry analyzer, Modular P800 Module, Roche Diagnostics Corp., Indpls., IN; GMMI No. 04998669001. Is a fully automatic, computerized chemistry analyzer intended for in vitro quantitative and qualitative determinations of a wide range analytes. The instrument is capable of performing potentiometric and photometric assays.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- Consignees were notified via an Urgent Medical Device Correction letter dated 11/12/08 and instructed to only execute "module unmasking" when the unit is in "standby" mode and to consult with their physicians at the facility to determine the clinical implications and whether any clinical intervention is needed for patients treated based upon results obtained with this software. Questions should be directed to Roche Diagnostics Technical Support at 1-800-428-2336.
- Quantity in commerce
- 22 systems in all
- Distribution
- Nationwide Distribution -- including states of California, Florida, Georgia, Hawaii, Indiana, Maryland, New York, North Carolina, Pennsylvania, and Virginia.
- Product codes and lots
- All units with software version 08-03.