Bio-Rad Variant II TURBO Hemoglobin Testing System with CDM v 4.0, model number 270-2600, Manufactured by...

Bio-Rad Laboratories Inc. / initiated 2008-10-17 / Terminated / root cause: labeling and UDI

Bio-Rad Laboratories Inc began a recall of Bio-Rad Variant II TURBO Hemoglobin Testing System with CDM v 4.0, model number 270-2600, Manufactured by Bio-Rad Laboratories, Hercules, CA Provides an... on . The reason FDA records is "Sample results may be assigned to the wrong barcode number and tube position numbers, leading to discrepant sample results". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1041-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
Bio-Rad Variant II TURBO Hemoglobin Testing System with CDM v 4.0, model number 270-2600, Manufactured by Bio-Rad Laboratories, Hercules, CA Provides an integrated method for sample preparation, separation and the percent determination of percent hemoglobin A 1 c in EDTA human whole blood. This device only tests for A 1 c.
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
Sample results may be assigned to the wrong barcode number and tube position numbers, leading to discrepant sample results.
Root cause tag
labeling and UDI
FDA root cause
Pending
Action
An Urgent Medical Device Correction letter was sent out on October 17, 2008 via fax and phone call. The letter recommended that customers review the Summary Report for duplicate tube position numbers or barcode IDs within the same rack. If there are questions or a need for assistance, customers are to contact their regional Bio-Rad office. Follow-ups will be made once root cause is identified.
Quantity in commerce
92 units
Distribution
Worldwide Distribution
Product codes and lots
version 4.0 software