Z-1009-2009

Roche COBAS c 111 Analyzer with ISE module option, GMMI No. 04777433001; Roche Diagnostics, Indianapolis, IN...

Roche Diagnostics GmbH / initiated 2008-11-17 / Terminated / root cause: manufacturing process

Roche Diagnostics Corp. began a recall of Roche COBAS c 111 Analyzer with ISE module option, GMMI No. 04777433001; Roche Diagnostics, Indianapolis, IN. Is an in-vitro diagnostic analyzer capable of... on . The reason FDA records is "Due to a manufacturing error at the supplier, not all cobas c 111 pinch valve caps are properly glued to the correct position on the pinch valve shaft. Without the glue, the cap can become loose or misaligned. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1009-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche COBAS c 111 Analyzer with ISE module option, GMMI No. 04777433001; Roche Diagnostics, Indianapolis, IN. Is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein electrolyte tests. Analytes are measured photometrically or turbidimetrically. The analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Due to a manufacturing error at the supplier, not all cobas c 111 pinch valve caps are properly glued to the correct position on the pinch valve shaft. Without the glue, the cap can become loose or misaligned. The effect of these faulty valves is that they may not open or close properly which can disturb the fluid transportation within the ISE module resulting in problems with ISE calibration, con
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Consignees were notified via an Urgent Medical Device Correction letter dated 11/17/08 and instructed Roche will contact them to make an appointment to inspect their analyzers. A work around was provided, for interim use, in the letter. Customers were instructed to complete and fax back the "ISE Controls Request - Purchased from a Distributor." A no charge replacement product will be shipped to customers.
Quantity in commerce
66
Distribution
Nationwide Distribution
Product codes and lots
Serial numbers: 50654, 50694, 50541, 50634, 50251, 50506, 50453, 50688, 50244, 50475, 50597, 50656, 50598, 50635, 50632, 50586, 50320, 50246, 50693, 50692, 50562, 50547, 50624, 50690, 50687, 50680, 50560, 50247, 50538, 50633, 50201, 50612, 50576, 50689, 50681, 50424, 50651, 50661, 50452, 50450, 50290, 50630, 50319, 50250, 50549, 50682, 50683, 50685, 50660, 50545, 50463, 50655, 50539, 50653, 50249, 50563, 50561, 50686, 50649, 50470, 50657, 50448, 50644, 50595, 50575, and 50679.