Z-0924-2009

Stryker PainPump Two Site Infusion Kit with ExFen 5.0", 120 mL Infusion Kit, sterile, 5 per box. Manufactured...

Stryker Corp. / initiated 2008-11-07 / Terminated / root cause: manufacturing process

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker PainPump Two Site Infusion Kit with ExFen 5.0", 120 mL Infusion Kit, sterile, 5 per box. Manufactured for Stryker Instruments, Kalamazoo, MI. Product... on . The reason FDA records is "There is a potential for the pump to overinfuse". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0924-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker PainPump Two Site Infusion Kit with ExFen 5.0", 120 mL Infusion Kit, sterile, 5 per box. Manufactured for Stryker Instruments, Kalamazoo, MI. Product number 0500-166-050, REF 500-166-50. A disposable syringe infuser which uses sustained vacuum pressure to deliver a continuous infusion of medications or fluids at a controlled rate. It delivers controlled amounts of medication directly to the intraoperative site for pain management and/or antibiotic administration. It is also intended for controlled delivery of local anesthetics in close proximity to nerves for postoperative regional anesthesia and pain management. Routes of administration may be intraoperative, perineural, or percutaneous.
Product code
FRN Pump, Infusion
Reason for recall
There is a potential for the pump to overinfuse.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
A "Medical Device Recall Notification" letter dated November 11, 2008 was issued to consignees. The letter identified the problem and instructed consignees to locate and destroy all affected products. Customers were also instructed to complete the enclosed Business Reply form and fax it to Stryker Instruments Regulatory Department at 866-521-2760. For questions regarding the recalled product, contact Stryker Instruments at 1-800-800-4236, extension 4056 or extension 3808.
Quantity in commerce
118 boxes of all products.
Distribution
Nationwide Distribution.
Product codes and lots
Lot Number: 2008011404 and 2008050201.