Z-0938-2009

VOYAGER NC Coronary Dilatation Catheter, 4.5 X 8.0 mm. The device is used for in the following: (1) Balloon...

Abbott Vascular / initiated 2008-11-05 / Terminated / root cause: packaging

Abbott Vascular-Cardiac Therapies dba Guidant Corporation began a recall of VOYAGER NC Coronary Dilatation Catheter, 4.5 X 8.0 mm. The device is used for in the following: (1) Balloon dilatation of the stenotic portion of a coronary... on . The reason FDA records is "The recall was initiated after Abbott Vascular received reports the protective, yellow balloon sheath of the 4.5mm diameter VOYAGER NC Coronary Dilatation Catheter has been tight. If unnoticed by the user, there is a...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0938-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Vascular
Product
VOYAGER NC Coronary Dilatation Catheter, 4.5 X 8.0 mm. The device is used for in the following: (1) Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion, (2) Balloon dilatation of a coronary artery occlusion for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction, and (3) Balloon dilatation of a stent after implantation.
Product code
MAF Stent, Coronary
Reason for recall
The recall was initiated after Abbott Vascular received reports the protective, yellow balloon sheath of the 4.5mm diameter VOYAGER NC Coronary Dilatation Catheter has been tight. If unnoticed by the user, there is a potential to cause damage to the balloon dilatation catheter during removal of the sheath. The balloon sheath covers the balloon, maintaining the fold configuration and protects the
Root cause tag
packaging
FDA root cause
Package design/selection
Action
An Urgent Device Recall letter (dated November 6, 2008) with an Efficacy Check Reconciliation form was issued and hand-carried to affected customers by their respective Sales Representative who assisted in locating and returning the recalled units. The letter described the problem and informed customers that Abbott Vascular will replace returned recalled product with like product or with VOYAGER NC sizes when they become available. Additionally, customers were instructed to contact their Abbott Vascular local account representative to review their inventory, complete and return the attached Efficacy Check Reconciliation Form and return affected products to Abbott Vascular.
Quantity in commerce
98 units
Distribution
Worldwide Distribution.
Product codes and lots
Part Number: 1011759-08, Lot Numbers: 8050961 and 8062361.