Z-0611-2009
ReSolve Hydrophilic Coated Locking Biliary Catheter, REF RBC-8-038, 8.5 F, Sterile EO, Merit Medical Systems...
Merit Medical Systems, Inc. / initiated 2008-11-07 / Terminated
Merit Medical Systems, Inc. began a recall of ReSolve Hydrophilic Coated Locking Biliary Catheter, REF RBC-8-038, 8.5 F, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. Used for drainage... on . The reason FDA records is "Biliary drainage catheter tips may separate from the catheter shaft after placement". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0611-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- ReSolve Hydrophilic Coated Locking Biliary Catheter, REF RBC-8-038, 8.5 F, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. Used for drainage of bile within the biliary system.
- Product code
- FGE Stents, Drains And Dilators For The Biliary Ducts
- Reason for recall
- Biliary drainage catheter tips may separate from the catheter shaft after placement.
- FDA root cause
- Other
- Action
- Customers were notified by Merit representatives by phone or visit beginning 11/07/2008. They were to identify and quarantine any unused inventory and complete a Product Retrieval Form. For further information, call 800-356-3748 x 4929.
- Quantity in commerce
- 284 units
- Distribution
- Nationwide. OUS distribution to Canada, Colombia, France, Germany, Mexico, Ireland, Netherlands, South Korea, and UK.
- Product codes and lots
- Lot numbers: G543418, MR604367, and MR607897.