ReSolve Hydrophilic Coated Locking Biliary Catheter, 12F, REF RBC-12-038, Sterile EO, Merit Medical Systems...

Merit Medical Systems, Inc. / initiated 2008-11-07 / Terminated

Merit Medical Systems, Inc. began a recall of ReSolve Hydrophilic Coated Locking Biliary Catheter, 12F, REF RBC-12-038, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. Used for drainage... on . The reason FDA records is "Biliary drainage catheter tips may separate from the catheter shaft after placement". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0613-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
ReSolve Hydrophilic Coated Locking Biliary Catheter, 12F, REF RBC-12-038, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. Used for drainage of bile within the biliary system.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Biliary drainage catheter tips may separate from the catheter shaft after placement.
FDA root cause
Other
Action
Customers were notified by Merit representatives by phone or visit beginning 11/07/2008. They were to identify and quarantine any unused inventory and complete a Product Retrieval Form. For further information, call 800-356-3748 x 4929.
Quantity in commerce
244 units
Distribution
Nationwide. OUS distribution to Canada, Colombia, France, Germany, Mexico, Ireland, Netherlands, South Korea, and UK.
Product codes and lots
Lot numbers: G543417, MR604365, and MR607896.