Z-1397-2009

Centricity TriWin Laboratory Information System; GE Healthcare Integrated IT Solutions, Barrington, IL 60010...

GE Healthcare / initiated 2008-11-17 / Terminated / root cause: software

GE Healthcare Integrated IT Solutions began a recall of Centricity TriWin Laboratory Information System; GE Healthcare Integrated IT Solutions, Barrington, IL 60010 The Centricity TriWin Laboratory product is a... on . The reason FDA records is "A software defect in the Centricity Lab (TriWin) product may incorrectly link patient information to a wrong report. Also, when viewing results in Patient Viewer, and a coagulation result is higher or lower than the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1397-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Centricity TriWin Laboratory Information System; GE Healthcare Integrated IT Solutions, Barrington, IL 60010 The Centricity TriWin Laboratory product is a laboratory information system. It is designed to facilitate the general clinical, anatomic pathology, and cytology laboratory workflow such as order entry, results entry, instrument interfacing, results reporting and patient record retrieval.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
A software defect in the Centricity Lab (TriWin) product may incorrectly link patient information to a wrong report. Also, when viewing results in Patient Viewer, and a coagulation result is higher or lower than the linearity limit, then all of the results following the higher or lower result will show the same result as the first flagged result.
Root cause tag
software
FDA root cause
Software design
Action
GE Healthcare notified their Centricity TriWin Laboratory customer via letter dated 11/17/08 of the software anomalies resulting in patient safety issues involving the patient information incorrectly being linked to a wrong report, and when viewing results in Patient Viewer, and a coagulation result is outside of a linearity limit, then all of the results following the out of linearity result will show the same result as the first flagged result. The user was informed that a software patch has been prepared to resolve the issues and will be installed on their system. Any questions were directed to GE Technical Support at 888-778-3375 or 905-479-1249, select option 1. The accounts was requested to complete the enclosed customer reply form, supplying the workstation software or build number, and fax it back to 905-305-0046 to schedule the software upgrade.
Quantity in commerce
26 units
Distribution
Worldwide Distribution -- USA and country of Canada.
Product codes and lots
software version 7.3