Abbott brand FreeStyle Navigator, Receiver Warranty Kit, Blood Glucose Monitoring System, Part Number...
Abbott Diabetes Care, Inc. began a recall of Abbott brand FreeStyle Navigator, Receiver Warranty Kit, Blood Glucose Monitoring System, Part Number: 70789-01. The product is manufactured by Benchmark... on . The reason FDA records is "The product has the Units of Measure set to mmol/L rather than mg/dL. The blood glucose results could be misinterpreted. This may lead to under treatment and potential for hyperglycemia". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1196-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Abbott brand FreeStyle Navigator, Receiver Warranty Kit, Blood Glucose Monitoring System, Part Number: 70789-01. The product is manufactured by Benchmark Electronics, 3535 Technology Drive NW, Rochester, MN and distributed by Abbott Diabetes Care, 1360 South Loop Road, Alameda, CA.
- Product code
- MDS Sensor, Glucose, Invasive
- Reason for recall
- The product has the Units of Measure set to mmol/L rather than mg/dL. The blood glucose results could be misinterpreted. This may lead to under treatment and potential for hyperglycemia.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- The Consignees of the affected FreeStyle Navigator Receiver Kit was contacted via telephone, Federal Express or an Abbott representative beginning on November 1, 2008. Contact Customer Service at 1-866-597-5520 with further questions.
- Quantity in commerce
- 2 units
- Distribution
- US Distribution to states of NY and AZ.
- Product codes and lots
- Lot Number: 01M2278; Serial Numbers: BAAH223-80915 and BAAH223-80182.