HR90K Platinum Series Headpiece Model Number AB-7300-001 ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar...

Advanced Bionics LLC / initiated 2008-07-24 / Terminated / root cause: manufacturing process

Advanced Bionics LLC began a recall of HR90K Platinum Series Headpiece Model Number AB-7300-001 ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The HiRes90K headpiece is used in... on . The reason FDA records is "Mistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0956-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Bionics LLC
Product
HR90K Platinum Series Headpiece Model Number AB-7300-001 ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The HiRes90K headpiece is used in conjunction with speech processors manufactured by Advanced Bionics. The Harmony HiResolution Bionic Ear System and Clarion Multi-Strategy Cochlear Implant are cochlear implant systems designed to provide useful hearing to individuals with severe-to-profound sensorineural hearing loss.
Product code
MCM Implant, Cochlear
Reason for recall
Mistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the headpiece's communication frequency is adjusted to match that of the internal device. Potential clinical symptoms include: -No lock -Intermittent lock -Implant locks, but cann
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Advanced Bionics is notifying clinicians of this situation so that they can provide appropriate clinical management. The notification letter sent to clinicians in the United States was sent via FedEx so that delivery information could be confirmed. The letter advises that Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Auria and Harmony headpieces are not affected by this issue. Tuning refers to the process through which the headpiece's communication frequency is adjusted to match that of the internal device. Since discovery of this issue firm has conducted an extensive technical analysis showing that the use of these headpieces does not pose any safety risk to the patient. While there has been no increase in the number of headpiece-related complaints, firm advises of the potential clinical symptoms: -No lock -Intermittent lock -Implant locks, but cannot measure impedances -Reduction in battery life, primarily for C1 patients with thick flaps. Per the letter, as long as the system maintains lock, there is no degradation in sound quality or stimulation. Further, the impact on battery life for CII and HiRes 90K patients is minimal. Medical professionals are advised that If they have any patients who display any of these clinical symptoms that they attribute to an affected headpiece, to advise the patient to contact Advanced Bionics Customer Service at 877-244-9541 and they will send them a replacement promptly.
Quantity in commerce
2,049 units
Distribution
Nationwide
Product codes and lots
Model #AB-7300-001; Serial Numbers: 167419, 167420, 169309, 169310, 169311, 170983, 170984, 170985, 172709, 172710, 174272, 174274, 174275, 174298, 175139, 175988, 175989, 175990, 175991, 177627, 177628, 177629, 177630, 177631, 177671, 177672, 177673, 178494, 179252, 181303, 181304, 183489, 184405 and 184694