Z-0980-2009

Cook Medical Universa Soft Ureteral Stent and Positioner with Hydrophilic Coating, 5.0Fr/26cm, sterile, Cook...

Cook Incorporated / initiated 2008-11-10 / Terminated / root cause: labeling and UDI

Vance Products Incorporated began a recall of Cook Medical Universa Soft Ureteral Stent and Positioner with Hydrophilic Coating, 5.0Fr/26cm, sterile, Cook Urological Inc., Spencer, IN; REF USH-526-R, GPN... on . The reason FDA records is ""Soft" stents were placed in "firm" stent outer boxes. The soft stent is only to be left in the patient for 6 months, but the firm stent can be left in the patient for 12 months, so this error could result in the stent...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0980-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook Medical Universa Soft Ureteral Stent and Positioner with Hydrophilic Coating, 5.0Fr/26cm, sterile, Cook Urological Inc., Spencer, IN; REF USH-526-R, GPN REF G49966. The product was placed in a shipping box which is preprinted with the word "firm" instead of "soft", but a product specific label stating "soft", as above, was also placed on the box. The ureteral stent is used for temporary drainage from the ureteropelvic junction to the bladder.
Product code
FAD Stent, Ureteral
Reason for recall
"Soft" stents were placed in "firm" stent outer boxes. The soft stent is only to be left in the patient for 6 months, but the firm stent can be left in the patient for 12 months, so this error could result in the stent being left in the patient for longer than indicated for the device.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Consignees were notified via recall letter, "Urgent: Recall Notice" dated November 10, 2008, and by follow-up phone call. Customers were instructed to isolate the recalled product, complete the enclosed form and fax to Cook Urological at 812-829-2353. After receipt of the completed form, Cook Urological will issue a Return Goods Authorization number for each customer to return the affected product to Cook Urological via FedEx or UPS.
Quantity in commerce
23
Distribution
Nationwide Distribution including states of Georgia, Oregon and Texas.
Product codes and lots
Lot Number: U1753668.