Z-0618-2009

Flexi-Seal Fecal Management System Advanced Odor Control Kit. Rx only; 1 silicone catheter tube assembly, 3-1...

ConvaTec Limited / initiated 2008-11-20 / Terminated / root cause: manufacturing process

ConvaTec began a recall of Flexi-Seal Fecal Management System Advanced Odor Control Kit. Rx only; 1 silicone catheter tube assembly, 3-1 Liter collection bags with filters, 1 luer-lock... on . The reason FDA records is "It was discovered that a portion of the inner nylon sleeve of the catheter was visible at the balloon end of a sample device". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0618-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConvaTec Limited
Product
Flexi-Seal Fecal Management System Advanced Odor Control Kit. Rx only; 1 silicone catheter tube assembly, 3-1 Liter collection bags with filters, 1 luer-lock syringe, SKU #411104 ConvaTec Skillman, NJ Made in USA with imported components "For the fecal management of patients with little or no bowel control and liquid or semi-liquid stool."
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
It was discovered that a portion of the inner nylon sleeve of the catheter was visible at the balloon end of a sample device.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Urgent Product Correction Letters were sent by first class mail on November 18, 2008 to Healthcare Professionals and ConvaTec Reps. The reason for the recall was identified along with possible risk to patient. Healthcare professionals were asked to use sufficient lubricating jelly during the insertion process with special care given to those who have a tendency to bleed or are on anti-coagulant medication. Physician should be notified if rectal bleeding or abdominal symptoms develop. ConvaTec reps were asked to contact all health care professionals to whom product or samples were given and reinforce the importance of proper lubrication and following the directions. A return form is to be completed by ConvaTec reps for each visit indicating the name of the Health Care Provider and that the they personally received notice of the product correction. All remaining sample products are to be returned to Besta, Franklin, WI.
Quantity in commerce
795
Distribution
Worldwide Distribution --- including USA and country of Canada.
Product codes and lots
Lot number 08-TM-51