Stryker Medical Stair-PRO Evacuation Chair civilian use stair chair, patient transport device, Model 6253
Stryker Medical Div. of Stryker Corporation began a recall of Stryker Medical Stair-PRO Evacuation Chair civilian use stair chair, patient transport device, Model 6253 on . The reason FDA records is "Handle could come off. The bolts may back out over time, resulting in a risk of dropping the patient and/or injury to transport personnel, a delay in medical treatment, or a delay in evacuation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0650-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Medical
- Product
- Stryker Medical Stair-PRO Evacuation Chair civilian use stair chair, patient transport device, Model 6253.
- Product code
- FPP Stretcher, Hand-Carried
- Reason for recall
- Handle could come off. The bolts may back out over time, resulting in a risk of dropping the patient and/or injury to transport personnel, a delay in medical treatment, or a delay in evacuation.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified by Urgent Medical Device Correction letter dated 11/18/08 and given the serial numbers of their units, an upgrade kit and instructions for performing replacement of the bolts. Consignees were informed Stryker will perform the upgrade if requested.
- Quantity in commerce
- 102
- Distribution
- Worldwide Distribution: USA, India, Netherlands and United Kingdom.
- Product codes and lots
- Serial numbers 080940895, 080940896, 080940578, 080940614, 080940615, 080940616, 080940618, 080940619, 080940620, 080940621, 080940623, 080940624, 080940625, 080940626, 080940628, 080940880, 080940571, 080940572, 080940573, 080940574, 080940575, 080940576, 080940629, 080940630, 080940599, 080940554, 080940555, 080940887, 080940888, 080940889, 080940890, 080940891, 080940892, 080940893, 080940894, 800940881, 080940772, 080940773, 080940774, 080940564, 080940565, 080940566, 080940850, 080940856, 080940556, 800940557, 080940558, 080940559, 080940560, 080940561, 800940562, 080940563, 080940633, 080940634, 080940635, 080940636, 080940245, 080940246, 080940851, 080940852, 080940853, 080940854, 080940846, 080940847, 080940848, 080940849, 080940810, 080940811, 080940812, 080940813, 080940814, 080940815, 080940577, 080940631, 080940632, 080940617, 080940622, 080940627, 080940818, 080940857, 080940882, 080940897, 080940816, 080940817, 080940567, 080940568, 080940569, 080940570, 080940579, 080940800, 080940801, 080940802, 080940878, 080940879, 080940855, 080940883, 080940884, 080940885, 080940886, 080940637, 080940638 and 080940803.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
